EVM-AC-03-03-02
EVOM™ Adjustable Warming Plate - Minimize the effect of temperature fluctuation by taking TEER measurements at the physiological temperature of 37°C.
- Overview
- Specifications
- Accessories
- Citations
- Related Products
Overview
There are 6 images available to view - click to enlarge and scroll through the product gallery.
EVOM™ Warming Plate (EVM-AC-03-03-02) Instruction Manual
/ Download as PDF
EVOM Warmer Data Sheet (EVM-AC-03-03-02)
/ Download as PDF
The Older Version Of EVOM™ Warming Plate (EVM-AC-03-03) Instruction Manual
/ Download as PDF
Go Beyond Temperature Monitoring–Actively Maintain Your Sample Plate at a Stable 37ºC for Optimal Results
The EVOM™ Warming Plate is a sample warming platform that actively maintains the physiological temperature of 37°C, ensuring stable and reliable TEER measurements when used with an EVOM™ Manual and an STX4 or STX HTS electrode. Unlike mere temperature monitoring, which only alerts users to fluctuations, maintaining the correct temperature prevents variability, enhances data accuracy, and preserves cell integrity. With warming parameters preset for users’ convenience, this system eliminates guesswork and provides a consistently controlled environment for optimal experimental results. Although Warming Plate is primarily for EVOM/TEER measurement, it also gives users flexibility for other applications which requires temperature range from 30°C to 45°C.
Features
- Compact 6, 12, 24, and 96 well plate warming platform
- Maintain sample temperature outside the incubator at 37°C
- Fast temperature equilibration/stabilization for TEER measurement
Benefits of EVOM™ Warming Plate
- Maintain the sample plate temperature at 37°C, the same as the incubator temperature, while the resistance/ TEER values are measured with WPI’s EVOM™ Manual and a pair of chopstick electrodes, such as STX4 or STX HTS.
- Heat a sample well plate from room temperature to 37°C in less than 12 minutes
- Minimize the effect of temperature fluctuation by taking measurements at the physiological temperature
- Save time by eliminating the need for room temperature equilibration before measurement
- Warm a sample well plate to the (incubator like) temperature of 37 °C
- Set plate warming at different temperatures between 30°C and 45°C
NOTE: The EVOM™ Warming Plate only has heating mechanism and can not cool. For accurate temperature the warming plates temperature set point should be >5°C above ambient room temperature
TEER readings are known to be affected by fluctuations in the sample temperatures. Normal practice has been to take the sample plates out of the incubator and let the plate equilibrate for 20 minutes at room temperature before measuring samples. WPI’s EVOM™ Warming Plate allows you to maintain sample plate temperature at 37°C, the same as the incubator temperature, while the resistance/TEER values are measured with WPI’s EVOM™ Manual and chopstick electrodes, such as STX4 or STX HTS. Using the EVOM™ Warming Plate, the effect of temperature fluctuation is minimized, and the measurement can be taken at the physiological temperature. This also saves your time by eliminating the requirement of room temperature equilibration before measurement. The EVOM™ Warming Plate platform will accept either a 6 well plate, 12 well plate, 24 well plate or 96 well plate.
Specifications
| Type | Description |
| Temp stability | ±0.5 °C |
| Display accuracy | 0.1 °C |
| Temperature range | programmable 30.0 °C to 45.0 °C* |
| Timing range | programmable 0–6 h 00 min |
| Heat up time | < 12 min from room temperature to reach set temperature of 37°C |
| Operating environment | Room temperature should be 20-25 °C and 35-50 % relative humidity |
| Dimension (L x W x H) | 17 x 16 x 12 cm |
| Power supply | 100-240 VAC, 50/60 Hz input to 24VDC-4A output |
| Power | 96 W |
| Weight | 1.6 Kg |
*The EVOM™ Warming Plate only has heating mechanism and can not cool. For accurate temperature the warming plates temperature set point should be >5°C above ambient room temperature