World Precision Instruments

Search the site by KEYWORD


Quick Finder


Testimonials

Researcher, University of Bristol

Our lab was interested in purchasing a syringe pump from WPI. The company's support was helpful in identifying the required model and even informed us of model-specific limitations (which were of no concern for our intended use). The overall processing and delivery was exceedingly fast – less tha...
read more

ISO-NOPNM

ISO-NOPNM

ISO-NOPNM Nitric Oxide Sensor - 100nm


Choose Quantity
  • Overview
  • Specifications
  • Accessories
  • Citations
  • Related Products

Overview

ISO-NOPNM

There are 1 images available to view - click to enlarge and scroll through the product gallery.

Instruction Manual
/ Download as PDF

  • For use with TBR4100 and TBR1025
  • Requires cable 91580 (sold separate)
  • Response Time: < 3s
  • Lowest Detection Limit: 0.5 nM
  • Sensitivity: 0.5 pA/nM
  • Application: Single Cell
  • Package of 3

Details

Measurement of NO at the cellular level

The ISO-NOPNM NanoSensor has a tip diameter of just 100nm and a detection limit for NO of less than 0.5nM — making it indisputably the smallest and most sensitive NO sensor in the world!

Design

The ISO-NOPNM is based on a novel design in which an electrochemically "activated" composite graphite nanofiber is used as the NO-sensing element. The surface of the Nanosensor is then modified using a unique multi-layered NO-selective membrane. Figure at right illustrates the response of the ISO-NOPNM following successive additions of nanomolar concentrations of NO. The ultra-low noise of the ISO-NOPNM (0.5pA) enables a detection limit of just 0.5nM NO. The response time of ISO-NOPNM is less than 3 seconds.

Selectivity of WPI's NO sensors

The ideal NO sensor should be insensitive to other reactive species likely to be present within the measurement environment. Conventional Nafion coated carbon fiber NO sensor exhibits a large response to such species. WPI's unique NO sensor technology utilizes an novel surface membrane which amplifies the response to NO whilst eliminating responses to a vast range of reactive species, including nitrite, ascorbic acid, hydrogen peroxide, catecolamines, and much more.

Specifications

Outside Diameter Conical Tip 100nm
Available Length 150µm
Response Time < 3s
Lowest Detection Limit/Range 0.5nM
Nominal Sensitivity-New sensor ≥0.5pA/nM
Baseline Drift none
Poise Voltage 865mV
Typical Quiescent Baseline Current, 25° 300pA
Acceptable Baseline Range 50-100pA
Polarization Time 1+ hours

Accessories

Citations

RelatedItems

Our Clients Include:

GlaxoSmithKline
University College London
Novartis
Imperial College
University of Cambridge
University of Oxford

Keep in Touch

We promise NEVER to share your details with anyone. You can opt out at any time.