General Purpose Probes (Type T Thermocouple)
A range of T Type thermocouple probes with options for immersion, surface or for air and gas temperature measurement.
Options
![]() |
HT-1 ''Workhorse'' General Purpose Probe - designed for temperature measurement of liquids, gases and semi-solids. Plastic handle with straight 3'' stainless shaft. Not good for surface temperatures. Maximum temperature 400°C, Time constant 0.5 Seconds. 5 foot lead. Not isolated. |
![]() |
HT-2 ''Workhorse'' General Purpose Probe - designed for temperature measurement of liquids, gases and semi-solids. Plastic handle with straight 9'' stainless shaft. Not good for surface temperatures. Maximum temperature 400°C, Time constant 0.5 Seconds. 5 foot lead. Not isolated. |
![]() |
BT-1 General Purpose Probe - Probably our best general purpose probe designed for temperature measurement of solids, liquids, gases and semi-solids. Welded stainless-stell shaft, plastic handle. Tip is .028'' diameter bent at right angle to probe to facilitate accurate surface measurements. Immersible. Maximum temperature 240°C, Time constant 0.15 Seconds. 5 foot lead. Not isolated. |
![]() |
DPT-12 Immersion Probe for Liquids - 12'' #304 stainless steel welded shaft. Wooden Handle. Diameter is .3/16'' diameter. Immersible. Maximum temperature 350°C, Time constant 2 Seconds. 5 foot lead. Not isolated. DPT-24 Immersion Probe for Liquids - 24'' #304 stainless steel welded shaft. Wooden Handle. Diameter is .3/16'' diameter. Immersible. Maximum temperature 350°C, Time constant 2 Seconds. 5 foot lead. Not isolated. |
![]() |
GT-1 Air and Gas Probe - 24'' #304 stainless steel shaft. Wooden Handle. For instant readings of temperature gradients in environmental chambers, freezer boxes, etc. Diameter is .3/16'' diameter. Immersible. Maximum temperature 200°C, Time constant Seconds. 5 foot lead. Not isolated. |
![]() |
PT-6 Sensor Six-Pack - For multi-point temperature measurements. Inexpensive, Flexible. Welded sensor beads, tough kapton insulation. Diameter is .029''. Maximum temperature 350°C, Time constant 0.01 Seconds. 5 foot lead. Not isolated. |
Time constant is defined as the time required to reach 63% of final temperature in liquid. An accurate reading is obtained when the reading stops changing, typically 5-10 time constants.
Maximum Temperature refers to intermit tent use of a probe. When probe is used for continuous measurements, maximum temperature should bereduced by 25%.
HT-1
General Purpose Type T Thermocouple Probe designed for temperature measurement of liquids, gases and semi-solids. Plastic handle with straight 3'' stainless shaft.
- 0.059" Tip diameter, 3" stainless steel probe length
- Temperature range -100°C to 400°C
- 0.5s time constant
- Not isolated
- 5' lead length
HT-2
General Purpose Type T Thermocouple Probe designed for temperature measurement of liquids, gases and semi-solids. Plastic handle with straight 9'' stainless shaft.
- 0.059" Tip diameter, 9" stainless steel probe length
- Temperature range -100°C to 400°C
- 0.5s time constant
- Not isolated
- 5' lead length
BT-1
General Purpose Type T Thermocouple Probe designed for surface temperature measurement of solids, liquids, gases and semi-solids. Welded stainless-stell shaft, plastic handle. Tip is bent at right angles to facilitate accurate surface measurements.
- 0.028" Tip diameter, bent at right angles, 4" proble length
- ±0.1°C in the physiological range
- Temperature range -50°C to 240°C
- 0.15s time constant
- Not isolated
- 5' lead length
Request
Catalogue
Chat
Print